基于磁共振成像与增强现实的计算机导航脊柱手术 Computer Navigated Spinal Surgery Using Magnetic Resonance Imaging and Augmented Reality 预印本

⚠️ 本文为 预印本(arXiv 公开版本),尚未经过正式同行评审。作者自发公开,结论未经期刊审稿验证,阅读时请知悉。
发布时间:2025-10-18
⬇️ 2 次下载
👁️ 预览文献
💡 AI 中文简介
当前脊柱疼痛治疗(如射频消融、硬膜外类固醇注射)依赖X光透视定位针头,会让患者和医生受到电离辐射。本研究开发了一套无辐射导航系统,把高分辨率磁共振扫描的腰椎模型与基于ArUco标记的增强现实技术结合,用ZED2立体相机实时追踪患者位姿,将MRI解剖图像叠加到患者体表上。在3D打印三椎体模型上的穿刺实验表明,双ArUco标记追踪比单标记更准确,平均穿刺偏差与传统透视技术约2毫米的偏差相当,有望替代透视成为更安全的脊柱导航方案。
领域医学图像引导手术 / 增强现实手术导航
适合谁读需要一定专业背景(医学影像、增强现实或手术导航领域)
核心贡献提出并验证了一套结合MRI与ArUco标记增强现实的无辐射脊柱手术导航系统;通过双标记追踪显著提高了穿刺精度,使平均偏差达到与传统透视技术相当的水平。
关键数据双ArUco标记追踪相比单标记提高了穿刺精度并降低了平均穿刺偏差;平均穿刺偏差与传统透视硬膜外技术约2毫米的平均偏差相当。
推荐理由想了解如何用增强现实和磁共振做出无辐射的脊柱穿刺导航,这篇工作给出了从建模到实验的完整验证,值得一读。
关键词["手术导航", "增强现实", "磁共振成像", "脊柱穿刺", "无辐射导航"]

原文摘要

查看英文原文摘要

Current spinal pain management procedures, such as radiofrequency ablation (RFA) and epidural steroid injection (ESI), rely on fluoroscopy for needle placement which exposes patients and physicians to ionizing radiation. In this paper, we investigate a radiation-free surgical navigation system for spinal pain management procedures that combines magnetic resonance imaging (MRI) with fiducial ArUco marker-based augmented reality (AR). High-resolution MRI scans of a lumbar spinal phantom were obtained and assembled as a surface mesh. Laplacian smoothing algorithms were then applied to smoothen the surface and improve the model fidelity. A commercially available stereo camera (ZED2) was used to track single or dual fiducial ArUco markers on the patient to determine the patient's real-time pose. Custom AR software was applied to overlay the MRI image onto the patient, allowing the physician to see not only the outer surface of the patient but also the complete anatomy of the patient below the surface. Needle-insertion trials on a 3D-printed 3-vertebra phantom showed that dual-ArUco marker tracking increased the accuracy of needle insertions and reduced the average needle misplacement distance compared to single-ArUco marker procedures. The average needle misplacement is comparable to the average deviation of 2 mm for conventional epidural techniques using fluoroscopy. Our radiation-free system demonstrates promise to serve as an alternative to fluoroscopy by improving image-guided spinal navigation.