近各向同性刚度与最小寄生误差的离轴柔顺远程运动中心关节 Off-Axis Compliant RCM Joint with Near-Isotropic Stiffness and Minimal Parasitic Error 预印本

⚠️ 本文为 预印本(arXiv 公开版本),尚未经过正式同行评审。作者自发公开,结论未经期刊审稿验证,阅读时请知悉。
发布时间:2026-03-30
👁️ 预览文献
💡 AI 中文简介
本研究设计了一种用于神经内镜操作的离轴柔顺远程运动中心关节,基于Tetra II概念将末端执行器置于四面体柔性结构外,以改善视线和消毒。通过两阶段设计优化刚度各向同性和约束面板,采用梁单元建模和有限元分析验证。PA12原型经选择性激光烧结制造,实验显示刚度椭圆主轴比1.37,寄生旋转比0.63%,在4.5°指令旋转下RCM漂移小于0.172毫米,疲劳预测工作空间12.1°-34.4°。实验与模拟趋势一致,偏差6-30%,为受限手术通道提供紧凑、可制造的RCM模块。
领域医疗机器人/手术机器人
适合谁读需要一定专业背景
核心贡献提出离轴单体制柔顺RCM关节设计,结合近各向同性刚度与最小寄生误差;通过两阶段优化方法平衡刚度各向同性与RCM漂移,并验证了可制造性。
关键数据刚度椭圆主轴比1.37;寄生旋转比0.63%;4.5°旋转下RCM漂移0.015-0.172 mm;疲劳预测工作空间12.1°-34.4°;实验与模拟偏差6-30%。
推荐理由该研究为受限手术通道提供了一种紧凑、可制造的RCM关节,实验验证了其低寄生误差和近各向同性刚度,对手术机器人设计有重要参考价值。
关键词["柔顺关节", "远程运动中心", "各向同性刚度", "寄生运动", "神经内镜", "有限元分析"]

原文摘要

查看英文原文摘要

This paper presents an off-axis, monolithic compliant Remote Center of Motion (RCM) joint for neuroendoscopic manipulation, combining near-isotropic stiffness with minimal parasitic motion. Based on the Tetra II concept, the end-effector is placed outside the tetrahedral flexure to improve line of sight, facilitate sterilization, and allow rapid tool release. Design proceeds in two stages: mobility panels are sized with a compliance-based isotropy objective, then constraining panels are synthesized through finite-element feasibility exploration to trade stiffness isotropy against RCM drift. The joint is modeled with beam elements and validated via detailed finite-element analyses, including fatigue-bounded stress constraints. A PA12 prototype is fabricated by selective laser sintering and characterized on a benchtop: a 2 N radial load is applied at the end-effector while a 6-DOF electromagnetic sensor records pose. The selected configuration produces a stiffness-ellipse principal axis ratio (PAR) of 1.37 and a parasitic-to-useful rotation ratio (PRR) of 0.63%. Under a 4.5° commanded rotation, the predicted RCM drift remains sub-millimetric (0.015-0.172 mm). Fatigue analysis predicts a usable rotational workspace of 12.1°-34.4° depending on direction. Experiments reproduce the simulated directional stiffness trend with typical deviations of 6-30%, demonstrating a compact, fabrication-ready RCM module for constrained surgical access.