内在标定扰动下椎骨DRR的投影域敏感性分析 Projection-Domain Sensitivity Analysis of Vertebral DRRs Under Intrinsic Calibration Perturbation 预印本

⚠️ 本文为 预印本(arXiv 公开版本),尚未经过正式同行评审。作者自发公开,结论未经期刊审稿验证,阅读时请知悉。
发布时间:2026-07-12
⬇️ 7 次下载
👁️ 预览文献
💡 AI 中文简介
本研究针对透视引导脊柱成像中几何标定精度对投影一致性的影响展开分析。通过CT椎骨模型和锥束成像几何,在保持解剖结构和采集位姿不变的情况下,生成具有真实和扰动内在标定参数的DRR图像,量化投影域变化。结果显示,即使微小的标定扰动也会引起可测量的投影几何、轮廓形态、标志点定位和DRR外观变化,且侧位投影比前后位投影更敏感。这些不一致会降低2D-3D配准精度,尤其是旋转对齐。研究提出投影域一致性可作为传统重建域标定指标的补充,为评估标定鲁棒性提供实用框架。
领域医学图像计算 / 图像引导介入
适合谁读需要一定专业背景,适合医学影像、图像引导介入领域的研究人员
核心贡献["首次系统分析内在标定扰动对椎骨DRR投影域一致性的影响,提出投影域一致性作为标定评估的补充指标", "建立合成框架,量化标定扰动对投影几何、轮廓、标志点及2D-3D配准的影响,并揭示视角依赖性"]
关键数据侧位投影比前后位投影表现出更大的轮廓变形和位移;标定扰动导致配准误差增加,尤其旋转对齐精度下降。
推荐理由如果你关心医学图像标定精度对配准的影响,这篇文章用严谨的合成实验给出了直观答案,值得一读。
关键词["2D-3D配准", "透视成像", "C臂标定", "数字重建放射影像", "椎骨成像", "投影域分析", "标定敏感性"]

原文摘要

查看英文原文摘要

Accurate geometric calibration is essential for fluoroscopy-guided spinal imaging, digitally reconstructed radiograph (DRR) generation, and 2D--3D vertebral registration. Although calibration quality is typically evaluated using reconstruction-based metrics such as reprojection error, its influence on projection-domain consistency remains poorly understood. This study presents a synthetic framework for evaluating how intrinsic calibration perturbations affect vertebral fluoroscopic projections and downstream registration performance. CT-derived vertebral models and controlled cone-beam imaging geometry were used to generate DRRs with both ground-truth and perturbed intrinsic calibration parameters while maintaining identical anatomy and acquisition pose. Projection-domain changes were quantified using anatomical landmark displacement, contour distance, silhouette overlap, image similarity, and landmark-based 2D--3D registration accuracy in anterior--posterior (AP) and lateral (LAT) views. Results show that even small intrinsic calibration perturbations produce measurable changes in vertebral projection geometry, contour morphology, landmark localization, and DRR appearance. Sensitivity is strongly view dependent, with LAT projections exhibiting substantially greater deformation and anatomical displacement than AP projections. These projection inconsistencies also degrade downstream 2D--3D registration, particularly rotational alignment accuracy. The findings demonstrate that projection-domain consistency complements conventional reconstruction-based calibration metrics and provides a practical framework for assessing calibration robustness. This approach may improve the reliability of DRR generation and fluoroscopy-guided vertebral registration in image-guided spinal applications.

📚 参考文献 (23 条)

查看23 条参考文献
  1. [1] Lin Li, Chaochao Zhou, Benjamin Albert, Junlin Guo, and Junchao Zhu. Rigid 2d–3d registration in medical imaging: A survey. Preprints, May 2026.
  2. [2] Lin Li, Benjamin Aubert, Paul Kemper, and Aric Plumley. Intrinsic tolerance in C-arm imaging: how extrinsic reoptimization preserves 3D reconstruction accuracy. In Maryam E. Rettmann and Pierre Jannin, editors, Medical Imaging 2026: Image-Guided Procedures, Robotic Interventions, and Modeling, volume 13927, page 139271T. International Society for Optics and Photonics, SPIE, 2026. 19
  3. Projection-Domain Sensitivity Analysis of Vertebral DRRs Under Intrinsic Calibration Perturbation A PREPRINT
  4. [3] J. B. Antoine Maintz and Max A. Viergever. A survey of medical image registration. Medical Image Analysis, 2(1):1–36, 1998.
  5. [4] Barbara Zitov´a and Jan Flusser. Image registration methods: a survey. Image and Vision Computing, 21(11):977– 1000, 2003.
  6. [5] P. Markelj, D. Tomaˇzeviˇc, B. Likar, and F. Pernuˇs. A review of 3d/2d registration methods for image-guided interventions. Medical Image Analysis, 16(3):642–661, 2012.
  7. [6] Junze Wang, Lei Fan, Dezheng Zhang, Weipeng Jing, Donglin Di, Yang Song, Sidong Liu, and Cong Cong. Visual prompt-agnostic evolution. In The Fourteenth International Conference on Learning Representations, 2026.
  8. [7] Graeme P. Penney, J. Weese, J. A. Little, P. Desmedt, D. L. G. Hill, and D. J. Hawkes. A comparison of similarity measures for use in 2-d/3-d medical image registration. IEEE Transactions on Medical Imaging, 17(4):586–595, 1998.
  9. [8] Paul Viola and William M. Wells. Alignment by maximization of mutual information. International Journal of Computer Vision, 24(2):137–154, 1997.
  10. [9] Yoshito Otake, Amy S. Wang, J. Webster Stayman, Jeffrey H. Siewerdsen, Ajoy J. Khanna, Jean-Paul Wolinsky, Ziya L. Gokaslan, Nicholas Theodore, and Gregory D. Hager. Automatic localization of vertebral levels in x- ray fluoroscopy using 3d-2d registration: a tool to reduce wrong-site surgery. Physics in Medicine & Biology, 57(17):5485–5508, 2012.
  11. [10] Yehyun Suh, Lin Li, Aric Plumley, Chaochao Zhou, Daniel Moyer, and Kongbin Kang. Pedicle screw pairing and registration for screw pose estimation from dual C-arm images using CAD models. In Maryam E. Rettmann and Pierre Jannin, editors, Medical Imaging 2026: Image-Guided Procedures, Robotic Interventions, and Modeling, volume 13927, page 139271W. International Society for Optics and Photonics, SPIE, 2026.
  12. [11] Nassir Navab, Michael Mitschke, and Alireza Bani-Hashemi. Camera augmented mobile c-arm (camc): calibra- tion, accuracy study, and clinical applications. IEEE Transactions on Medical Imaging, 18(9):806–816, 1999.
  13. [12] Jeffrey H. Siewerdsen. Cone-beam ct with a flat-panel detector: from image science to image-guided surgery. Nuclear Instruments and Methods in Physics Research Section A, 648(Supplement 1):S241–S250, 2011.
  14. [13] Wojciech Zbijewski et al. Volume-of-change cone-beam ct for image-guided surgery. Physics in Medicine & Biology, 56(21):6855–6871, 2011.
  15. [14] Vanessa Su and Xiaofeng Yang. KAN-DDPM: Kolmogorov-Arnold networks with diffusion denoising proba- bilistic models for MRI-to-CT synthesis. In Shandong Wu, editor, Medical Imaging 2025: Imaging Informatics, volume 13411, page 1341110. International Society for Optics and Photonics, SPIE, 2025.
  16. [15] Robert L. Siddon. Fast calculation of the exact radiological path for a three-dimensional ct array. In Medical Physics, volume 12, pages 252–255, 1985.
  17. [16] Andreas Maier et al. Precision, accuracy and performance of gpu-based rigid multi-modal image registration. Medical Physics, 40(9):091907, 2013.
  18. [17] Vignesh Gopalakrishnan et al. Differentiable x-ray rendering and 2d–3d registration for medical imaging. Med- ical Image Analysis, 2022.
  19. [18] Shun Miao, Z. Jane Wang, and Rui Liao. A cnn regression approach for real-time 2d/3d registration. IEEE Transactions on Medical Imaging, 35(5):1352–1363, 2016.
  20. [19] Mathias Unberath et al. Deepdrr – a catalyst for machine learning in fluoroscopy-guided procedures. Interna- tional Conference on Medical Image Computing and Computer-Assisted Intervention, 2018.
  21. [20] Vanessa Su, Xiaohan Yuan, Shaoyan Pan, Mingzhe Hu, Xiaoqian Chen, Sibo Tian, and Xiaofeng Yang. TTT- SEG: a multi-structure segmentation on cardiac MRI using vision testing-time training network. In Barjor S. Gimi and Andrzej Krol, editors, Medical Imaging 2025: Clinical and Biomedical Imaging, volume 13410, page 134101X. International Society for Optics and Photonics, SPIE, 2025.
  22. [21] Junze Wang, Lei Fan, Weipeng Jing, Donglin Di, Yang Song, Sidong Liu, and Cong Cong. Hypergraph tversky- aware domain incremental learning for brain tumor segmentation with missing modalities. In International Conference on Medical Image Computing and Computer-Assisted Intervention, pages 283–293. Springer, 2025.
  23. [22] Chaochao Zhou, Thomas Cha, Yun Peng, and Guoan Li. Transfer learning from an artificial radiograph-landmark dataset for registration of the anatomic skull model to dual fluoroscopic x-ray images. Computers in Biology and Medicine, 138:104923, 2021. 20