面向高精度机器人针引导的物理约束运动学标定 Physically Constrained Kinematic Calibration for High-Precision Robotic Needle Guidance 预印本

⚠️ 本文为 预印本(arXiv 公开版本),尚未经过正式同行评审。作者自发公开,结论未经期刊审稿验证,阅读时请知悉。
发布时间:2026-10-30
⬇️ 1 次下载
👁️ 预览文献
💡 AI 中文简介
本研究针对6自由度机器人针引导 manipulator,提出一种带物理边界约束的运动学标定方法,用合理的参数范围作为优化约束来抑制参数耦合与过拟合,从而稳定识别19个几何参数。仿真中末端均方根误差从3.422毫米降至0.690毫米,提升79.8%;硬件实验验证精度达亚毫米级,轨迹验证和坐标格点验证的均方根误差分别为0.865毫米和0.610毫米。该方法为临床高精度针靶向奠定了基础。
领域医疗机器人/机器人运动学标定
适合谁读需要一定专业背景(机器人学、优化或医疗机器人领域读者)
核心贡献1. 将物理合理的参数范围作为主动优化约束而非数值保护,有效缓解参数耦合与过拟合,实现高度稳定的几何参数辨识;2. 在6自由度针引导机器人上验证了亚毫米级标定精度,为临床针靶向提供高精度基础。
关键数据仿真末端RMSE从3.422 mm降至0.690 mm(提升79.8%);硬件实验轨迹验证RMSE 0.865 mm,坐标格点验证RMSE 0.610 mm。
推荐理由如果你关心手术机器人如何把毫米级误差压到亚毫米,这篇用物理约束做标定的工作值得一读。
关键词["运动学标定", "机器人针引导", "物理约束优化", "参数辨识", "经皮穿刺", "手术机器人"]

原文摘要

查看英文原文摘要

Percutaneous needle insertion is a common practice in minimally invasive procedures whose clinical success depends on accurate needle placement. However, achieving the necessary precision is a challenge for manual operators. Robotic assistance can improve targeting accuracy and reduce operator dependence, but residual millimeter-scale errors persist because nominal kinematic models do not fully represent the physical manipulator. This paper presents a robust bound-constrained kinematic calibration method designed to improve positional accuracy for a 6-Degree-of-Freedom (6-DoF) robotic needle-guidance manipulator. The approach uses physically plausible parameter ranges to regularize a compact 19-parameter geometric identification model. By treating these bounds as active optimization constraints rather than mere numerical safeguards, the framework effectively mitigates parameter coupling and overfitting, ensuring highly stable parameter identification. In simulation, the proposed method reduced end-effector Root Mean Square Error (RMSE) from 3.422 mm to 0.690 mm (a 79.8% improvement). In benchtop hardware experiments, the calibrated model achieved sub-millimeter validation accuracy, including 0.865 mm RMSE in experimental trajectory validation and 0.610 mm RMSE in a coordinate-lattice verification task. These results demonstrate that physically constrained kinematic calibration can enhance robotic needle-targeting accuracy, establishing a high-precision foundation for future clinical applications.